Surgical Cautery
Logic - 100dx
|
 |
Logic-100dx
Electro-Surgical unit, which is suitable for microsurgery as well as
for precision monopolar and bipolar surgery. Logic-100dx can supply RF
operative power carried out by a hand-piece or a footswitch.
• User Friendly
• Enhanced safety
• A New Procedure technology
• Light weight and compact design
• Safe, Precise
• For micro and minor procedure economical
In the uses of surgical diathermy, the technical ability of the unit
combined with plain safety is of prime importance-both for the patient
and the user. There must be no possibility of accidental, undesirable
results in the use of LevelUP Electro-Surgical Units been
developed.
The LevelUP Logic 100 DX
are newly developed Electro-Surgical units of particular safety and
reliability. The Logic 100 DX as the most refined unit of its kind in
the world.
Advanced Performance Features
Monopolar
Cut Modes:
• Pure Cut: Suitable for Micro and Minor procedure.
•
Blend-I: Preset blend I option for cutting with change in degree of
hemostasis suited for Gyne procedures.
• Blend-II: Preset blend II mode allows for cutting and simultaneous coagulation.
Three Coag Modes:
• Soft Mode: For low voltage coagulation, suitable for laparoscopic and delicate tissue work.
•
Forced Mode: This operating mode covers all the demands of standard
coagulation. The surgeon can work quickly and effectively.
• Spray: For non-contact large tissue area coagulation with minimum depth of necrosis.
Bipolar
Two Modes:
•
Micro: For delicate tissue work by control desiccation without
charring, adhesion or blancing of adjacent tissue (ophthalmic &
neuro-surgery).
•
Coag: High current low voltage pulsed bipolar output for high-pressure
vessel sealing with ligation sure artery forceps. Also suitable for
bipolar needle ablation.
Advanced Safety Features
• Patient Electrode Monitoring Technology
monitors the quality and quantity of contact between the patient and
the patient plate. If this contact is not proper the unit deactivates
and alarms, thus eliminating risk of passive electrode site burns. It
also deactivates the unit and alarms in case of any discontinuity in
the patient plate cord.
• Isolated RF output
directs current to return back to the generator to reduce the risk of
alternative site burns and to ensure patient safety in the case of any
discontinuity in patient plate cord. This current has no reference to
ground so it does not cause noticeable interference with the ECG
monitor, Cardiac pacemaker etc.
• Independent activation
of unipolar and bipolar output to protect against inadvertent output
from an idle accessory lying on or near the patient or surgical staff.
• Class "2" leakage cancelling, a must for lengthy procedures like Cardiac surgery (CABG) etc.
• Ergonomically designed trolley to house cautery and accessories.
• Small and light weight portable unit.
• Smart Modular Technology for fast and effective after sales support (service).
WEIGHT & DIMENSION
Width : 307 mm
Height : 124 mm
Depth : 323 mm
Weight : 6 Kgs.
OUTPUT WAVEFORMS
Bipolar : 330 KHz sinusoid
Monopolar Cut
Pure : 330 KHz sinusoid
Blend I : 330 KHz sinusoid
Blend II : 330 KHz sinusoid
Monopolar Coag.
Desiccate : 330 KHz Pulse
Fulgurate : 330 KHz Pulse
LOW FREQUENCY LEAKAGE (50-60 Hz)
Source current, patient leads, all outputs tied together.
| • |
Normal polarity, intact
chassis ground |
< 10 mA |
| • |
Normal polarity, ground open |
< 50 mA |
| • |
Reverse polarity, ground open |
< 50 mA |
| • |
Sink current, 140V, applied,
all inputs |
< 50 mA |
|
|
INPUT POWER REQUIREMENTS
Operating range is 180 – 260 volts. Maximum current is 2.5 amperes (in cut, 240V nominal).
HIGH FREQUENCY LEAKAGE
Bipolar : Less than 40 mA rms
Monopolar : Less than 70 mA rms
POWER READOUTS
Agree with actual power into rated load to within + 15% or 5 watts, whichever is greater. Dosage error monitor.
COOLING
Convection, no fan
AUDIO VOLUME
The mode indicator tones are adjustable to a minimum level of 45 dB at
1 meter. The alarm tones are not adjustable and are set at 65 dB
minimum at 1 meter.
|
CUT
|
MACRO
|
MICRO
|
Crest Factor
|
|
MONOPOLAR CUT
BLEND I
BLEND II
|
100 w / 500
Ohms
70 w / 300
Ohms
50 w / 300
Ohms
|
30 w / 500
Ohms
30 w / 300
Ohms
30 w / 300
Ohms
|
1.5
2.5
3.0
|
|
COAG.
|
|
|
|
|
SOFT
FORCED
SPRAY
|
50 w / 300
Ohms
70 w / 300
Ohms
50 w / 500
Ohms
|
30 w / 300
Ohms
30 w / 300
Ohms
30 w / 500
Ohms
|
5.0
5.0
8.0
|
|
BIPOLAR /
COAG.
|
|
|
|
|
BIPOLAR
|
50 w / 100
Ohms
|
30 w / 100
Ohms
|
5.0
|
|